Feedback Mode
Feedback Mode measures every compression and breath on each manikin and shows the assessment on its card right away. Green means within the target range and red means outside it, so students can watch the screen and correct themselves immediately.
Requirements
Section titled “Requirements”Overview
Section titled “Overview”- Start
- Measure
- Pause
- Finish
- Saved automatically
- Score cards
- Result details
How to use
Section titled “How to use”Ready screen
Section titled “Ready screen”Each connected manikin has one card. The large colored circle is the manikin ID. Cards of disconnected manikins are dimmed.
- 1Connected count: How many of the selected manikins are connected. Example: 4/4 connected
- 2Select view: Shows just one manikin large, or goes back to All. Shown only when there are 2 or more manikins.
- 3Settings (gear icon): Opens the Quick Settings window (step 2).
- 4Start Assessment: Starts the measurement (step 3).
- 5Search: Finds and connects more manikins (Manikin connection).
Quick Settings
Section titled “Quick Settings”- 1CPR Measurement Items: Hands-only CPR (compressions only) or Conventional CPR (compressions + breaths). The default is Hands-only CPR.
- 2Compression Rate Guide: Turns the beat sound on or off and sets Speed to 100, 110, or 120 (times per minute). The default is on, 110. The sound plays only during measurement.
- 3Layout: The number of rows and columns of cards. Auto or Fixed (you set the rows and columns). With Fixed, if rows × columns is less than the number of manikins, the cards are arranged automatically. For details, see Display Settings.
- 4More settings: Goes to the Settings screen.
- 5Changes apply only when you select Done. Select Cancel to discard your changes.
Measurement start
Section titled “Measurement start”Select Start Assessment to start measuring. The cards update with every compression.
- 1Clock button: Shown in place of the start button and shows the elapsed time (min:sec). Select it to pause (step 4).
- 2Flip all: Turns all visible cards to the front or to the back at once. Shown only during measurement.
- 3Card border color: Green = Good, red = needs correction, gray = no compressions yet. Good means the depth, rate, and recoil of the last compression are all within the target range (during breaths, the breath volume). For the ranges, see Target ranges below.
- 4Card flip: Select a card to flip only that card and show its back, with large text.
During measurement, the Home and Settings buttons are grayed out and the Search button is not shown.
Measurement end
Section titled “Measurement end”Select the clock button to pause the measurement and open this window.
- 1Resume: Continues the measurement.
- 2Quit: Ends the measurement and saves it automatically.
Score cards
Section titled “Score cards”- 1Hearts · total score (%): 3 hearts = 75% or more, 2 hearts = 50% or more, and 1 heart otherwise.
- 2Medals: The cards with the top 3 total scores get gold, silver, and bronze medals.
- 3Numbers at the bottom: Compression icon = compression score, breath icon = breath score (only when breaths were measured), clock icon = points lost to hands-off time (time with hands off the chest). Lost points are shown as a negative number, and the value is 0 if hands-off time is not included in the score. The box below explains how it is calculated. Select a card to go to Result details.
- 4Cards showing “No results.” are dimmed and cannot be opened.
Select Start Assessment again to start a new measurement. You can review past records in the Results list.
Measurement card
Section titled “Measurement card”The front of a card has the rate gauge above the manikin picture and the depth bar next to it. The color number at the top left is the manikin ID, and the face icon next to it is the manikin’s target patient. The target patient is set for each manikin type, and you can change it in Assessment Settings.
During compressions
Section titled “During compressions”- 1One-word feedback: Based on the last single compression. If depth, rate, and recoil are all within the target range, it shows Good (green). If something is off, one message is shown in red, in the order of what to correct first (Recoil needed · Weak · Strong · Fast · Slow). During breaths, it is based on the breath volume. Before the first compression, Start compressions is shown in gray.
- 2Rate gauge: Compressions per minute. The number in the center is the current rate. The gauge is green when the needle is in the dark gray band at the top (100–120 per minute) and red outside it.
- 3Hand position: The dot on the chest shows your hand position and is briefly marked green (center) or red (off center) at each compression. It works only on manikins with a position sensor and is not part of the one-word feedback. If you need a position sensor, please contact contact@imlabworld.com.
- 4Depth bar: Shows compression depth from top to bottom. The dark gray band is the right depth, and the colored line is the deepest point of this compression (green if right). The section at the very top is green if the depth was right and recoil was complete, and red otherwise.
- 5Good / total compressions: The number of good compressions (green) and all compressions in the current cycle.
In Hands-only CPR, when compressions stop, the number in the center changes to the time with hands off the chest (seconds).
During breaths (Conventional CPR)
Section titled “During breaths (Conventional CPR)”- 1Status icon: A breath icon during breaths, and a compression icon during compressions.
- 2Breath timer: Shows the time since compressions stopped, up to 12 seconds. The section where a breath was given is shown in green (right) or red.
- 3Number in the center: The time with hands off the chest (seconds). After 10 seconds, it is shown in red.
- 4Breath volume bar: Fills from bottom to top. The dark gray band is the right breath volume.
Card back
Section titled “Card back”Select a card or the Flip all button to show the back of the card.
- 1One-word feedback: Shown large. It follows the same rules as ① in step 1.
- 2Numbers: During compressions, depth (cm) · rate (CPM). During breaths, breath volume (mL) · time with hands off (s). In Hands-only CPR, when compressions stop, only the time with hands off is shown. Depth, rate, and breath volume are green within the target range and red outside it. The time with hands off is gray, and turns red after 10 seconds.
Target ranges
Section titled “Target ranges”These are based on the default guideline AHA 2025 and an adult manikin. Values outside the range are shown in red, and pressing deeper than the range is red too.
| Item | Where it is shown | Target range |
|---|---|---|
| Compression depth | Depth bar | 5–6 cm |
| Recoil | Section at the top of the bar | Depth is right, and the chest comes back up to within 1 cm |
| Compression rate | Rate gauge needle · number in the center | 100–120 per minute |
| Hand position (with a position sensor) | Dot on the chest | Center |
| Breath volume | Breath volume bar (during breaths) | 400–600 mL |
| Hands-off time | Number in the center (seconds) | Under 10 seconds |
When viewing on a student tablet
Section titled “When viewing on a student tablet”Frequently asked questions
Section titled “Frequently asked questions”I can’t select Start Assessment
When no manikins are connected, Start Assessment is grayed out. Connect manikins with the Search button at the bottom right. A measurement cannot be started on a student tablet.
It says “No manikin connected.”
It means no manikins are connected to Feedback Mode. Select the Search button at the bottom right to open the manikin search screen, then select and connect manikins to show their cards. For details, see Manikin connection.
I selected the clock button right after starting, but nothing happens
The pause window does not open until 1 second after the start. Wait a moment and select it again.
I can’t go back to Home during measurement
The Home button is grayed out during measurement. Select the clock button, then select Quit to end the measurement before you leave.
The card doesn’t flip when I select it
Cards flip only during measurement. They do not flip on the ready screen or the score card screen. Start the measurement, then select the card.
I can’t see the Select view button
The Select view button is shown only when 2 or more manikins are selected on the manikin search screen. Select and connect 2 or more manikins.
Selecting a score card doesn’t open the result
If no compressions or breaths are measured on that manikin during the measurement, no result is saved. That is why the card shows “No results.” and cannot be opened. Next time, start the measurement first, then press that manikin’s chest.
There’s no beat sound
Check that the Compression Rate Guide switch in Quick Settings is on. The sound plays only during measurement. Also check the PC speaker volume.
I selected Conventional CPR, but breaths aren’t measured
A manikin without the part that measures breaths (breath module) measures only compressions, even if you select Conventional CPR. Check whether your manikin has a breath module.