Set the standards the app uses to judge and score students’ compressions and breaths. The compression depth and breath volume ranges depend on the guideline and the target patient (adult, child, or infant). You also set here which items count toward the score.

Select the gear Settings button at the top right of the Home screen, then select Assessment Settings. In Feedback Mode, selecting More settings in the Quick Settings window also opens the Settings screen. (Feedback Mode)

1

The value currently selected is shown in parentheses next to each item name.

Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsReset12345
  1. 1Reset: Returns all assessment settings to their defaults. (step 6)
  2. 2Guideline: Selects the guideline used as the scoring standard. (step 2)
  3. 3Compression Rate Guide: Selects the metronome speed (BPM). (step 3)
  4. 4Target Patient Mode: Selects the target patient for each manikin type. (step 4)
  5. 5Scoring / CPR Mode: Sets the CPR mode and the items that count toward the score. (step 5)
2
Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsResetGuidelineAHA 2025Heart & Stroke 2025Canadian Red Cross 2025KACPR 2025ANZCOR(ARC) 2026SRFAC 2021ERC 2025CancelConfirm12
  1. 1Select the guideline you want to use. The default is AHA 2025.
  2. 2Your choice is saved only when you select Confirm. If you select Cancel, nothing changes.

Normal compression depth ranges by guideline and target patient:

Guideline Adult Child Infant
AHA 2025 · Heart & Stroke 2025 · ERC 2025 5–6 cm At least 5 cm At least 4 cm
Canadian Red Cross 2025 5–6 cm At least 5 cm At least 4 cm
KACPR 2025 5–6 cm At least 4 cm At least 4 cm
ANZCOR(ARC) 2026 5–7 cm At least 5 cm At least 4 cm
SRFAC 2021 4–6 cm At least 4 cm At least 3 cm

Compressions that are too deep are also shown in red.

Normal breath volume ranges when breaths are turned on:

Guideline Adult · Child Infant
All except SRFAC 2021 400–600 mL 50–70 mL
SRFAC 2021 400–600 mL 30–50 mL
3
Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsResetCompression Rate Guide100 BPM110 BPM120 BPMCancelConfirm12
  1. 1Select 100, 110, or 120 BPM. The default is 110 BPM.
  2. 2Your choice is saved only when you select Confirm.

This is the beat speed that plays in Feedback Mode when the metronome is on. The scoring standard (100–120 per minute) does not change.

This is the same setting as Compression Rate Guide → Speed in the Feedback Mode Quick Settings window. If you change it in one place, it changes in the other too.

4

When expanded, the manikin types currently connected are shown, one per row.

Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)Basic Buddy+(ADULT)Little Junior(CHILD)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsReset123
  1. 1Select the Target Patient Mode row to expand or collapse it.
  2. 2Select the manikin type you want to change. All manikins of the same type change together.
  3. 3Current target patient: ADULT = Adult, CHILD = Child, INFANT = Infant.
Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)Basic Buddy+(ADULT)Little Junior(CHILD)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsResetTarget Patient SetupAdultChildInfantCancelConfirm123
  1. 1Select Adult, Child, or Infant.
  2. 2Grayed-out options are target patients the manikin does not support, so you cannot select them.
  3. 3Your choice is saved only when you select Confirm.

Target patients you can select for each manikin, and their defaults:

Manikin (name shown in the app) Target patients you can select Default
Basic Buddy+ · BASICBilly+ · cprCUBE PRO · CPR Prompt+ Adult, Child Adult
Ambu Junior · Little Junior Child Child
Ambu Baby · Baby Buddy+ · Prompt Infant+ · Jana · Little Baby Infant Infant
Other manikins (Little Anne, Brayden, etc.) Adult Adult
5
Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Conventional CPR)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsReset12345
  1. 1Breath: Sets the CPR mode. Off means Hands-only CPR (compressions only), and on means Conventional CPR (compressions + rescue breathing). It is off by default. Compression just above it is always on and cannot be changed.
  2. 2Compression:Breath Score Weight: Drag the slider to set how much compressions and breaths each count toward the score. It adjusts in 5% steps, between 5% and 95% for compressions. It is grayed out when breaths are off.
  3. 3If you turn off Rate or Recoil, that item does not count toward the compression score. Depth is always on and cannot be changed.
  4. 4You cannot change Breath Volume directly. It turns on and off with the Breath switch above.
  5. 5Hands-off Time: When on, the longer compressions stop, the more points are deducted from the total score. When off, no points are deducted.

Example: if you measure Conventional CPR with the defaults (compression 75% / breath 25%), the score card shows compressions out of “/75” and breaths out of “/25”.

6

Select Reset at the top right to show a confirmation window.

Guideline(AHA 2025)Compression Rate Guide(110 BPM)Target Patient ModeScoring / CPR Mode(Hands-only CPR)CPR Measurement ItemsCompression:Breath Score WeightCompression Score ItemsBreath Score ItemsHands-off Time ItemsCompressionBreathDepthRateRecoilBreath VolumeHands-off TimeCompression75%Breath25%Assessment SettingsResetReset Assessment SettingsRestore assessment settings to defaults?Current settings will be lost.CancelReset1
  1. 1Select Reset to return to the defaults below. If you select Cancel, nothing changes.
Item Default
Guideline AHA 2025
Compression Rate Guide 110 BPM
Target Patient Mode Default for each manikin type (table above)
CPR Mode Hands-only CPR (breaths off)
Compression:Breath Score Weight Compression 75% / Breath 25%
Rate · Recoil · Hands-off Time All on
Feedback Mode metronome (Compression Rate Guide switch) On
Mission Mode Session Time Limit 1 min
Mission Mode “End the session when the first person reaches 100%” Off
I selected Target Patient Mode and it says “No manikin connected.”

The target patient is set for each connected manikin type, so it cannot be set when no manikin is connected. Connect a manikin first, then select it again. (Manikin connection)

Child (or Infant) is grayed out and I can’t select it

The connected manikin does not support that target patient. You can select child only when Basic Buddy+, BASICBilly+, cprCUBE PRO, CPR Prompt+, Ambu Junior, or Little Junior is connected, and infant only when Ambu Baby, Baby Buddy+, Prompt Infant+, Jana, or Little Baby is connected. The full list by manikin is in the “Target patients you can select” table above.

The score weight slider won’t move

When the Breath switch is off (Hands-only CPR), the slider is grayed out. Turn on breaths to adjust it.

If I change the Compression Rate Guide, does the scoring standard change too?

No, only the metronome beat speed changes. The compression rate standard is always 100–120 per minute.

I switched to Conventional CPR, but there’s no breath score

Manikins that cannot measure breaths are scored on compressions only. Check whether the manikin has a breath module.

Settings won’t open on the student tablet

The settings screen does not open on a student tablet. Change settings on the instructor PC (or instructor tablet).